In December 2021, imtmedical ag, a subsidiary of Vyaire Medical, issued a bellavista ventilator voluntary correction due to risk the ventilators can cease ventilation, according to an FDA alert.

According to the FDA, with specific software version and configuration combinations, the ventilators can cease ventilation and generate a technical failure alarm 305.

There is a potential issue with bellavista 1000 and 1000e ventilators that have software version 6.0.1600.0 or higher installed, which was deployed in February 2021, and also have the Data Communication port set to “HL7,” the FDA reports.

Customers with software version 6.0.1600.0 or higher installed have been notified by Vyaire to immediately disable the HL7 data communication if activated. If the HL7 data communication is not disabled, such devices are at risk of the reported unintentional ceasing of ventilation, according to the FDA.

This decision is based on the receipt of reports that some bellavista 1000 and 1000e ventilators have unintentionally ceased ventilation during clinical use and require rebooting to resume ventilation due to a software issue. The FDA reports that an internal investigation determined that bellavista 1000 and 1000e ventilators with software version 6.0.1600.0 or higher installed can have a conflict in memory resource allocation between software tasks when the Data Communication port is set to “HL7.”

According to Vyaire, no patient harm has been reported related to this issue. If clinicians operate a bellavista ventilator and do not set Data Communication port to “HL7”, this error will not occur. “Therefore, Vyaire supports the continued use of these devices and directs customers to follow the instructions provided in the Company’s correction notification to immediately disable the HL7 data communication if activated while the company develops and deploys a software update to address the issue, anticipated in Q1 of calendar year 2022.”

The bellavista ventilator voluntary correction includes ventilators that were distributed to medical facilities in the US between 3/21/19 through 12/23/21 located in the following states: AL, CA, CO, CT, DC, FL, GA, ID, IL, IN, LA, MA, MD, MI, MO, MS, NC, NJ, NY, OH, OK, PA, PR SD, TX, VA, WA, WI and WV.